your best partner for vacuum

AP29ECO Probe

AP29ECO Sampling Probe
AP29ECO, an accessory to Sensistor ISH2000 Hydrogen Leak Detector, allows you to do automatic leak testing with hydrogen tracer gas. Examples of AP29ECO Sampling Probes techniques are integral test in accumulation chamber, clam shell test, direct point sniffing and scanning. Controlled by the Sensistor ISH2000 Hydrogen Leak Detector, it draws a well-defined sample of air and passes it over the built-in hydrogen sensor. It also handles calibration of your entire leak test system. The sampling probe has an automatic purging function which can be activated before and after sampling, and when the gas concentration exceeds a set limit. It can, therefore, encounter gross leaks and still be cleared within seconds. Together with the unique properties of hydrogen tracer gas, this feature gives an unprecedented reliability of your system.

The AP29ECO is built for heavy-duty applications. Its rugged design and serviceability make it ideal for use in tough industrial environments. The sensor element can be replaced without opening the probe, and the sniffer flow function is powered by a long-life membrane pump. The pump is easy to maintain and it is only in operation when a sample is drawn. This makes the AP29ECO a cost effective and environmentally friendly alternative. The unit can be ordered with two different sniffer flows — 1 cc/s or 3 cc/s.

Benefits

  • High reliability for demanding industrial applications
  • For semi- or fully automatic leak testing
  • Easy to maintain
  • Easy sensor change
  • Overexposure protection minimizes recovery time in case of gross leaks
  • Automatic calibration of leak test system

Features

  • Sniffer flow alarm
  • The AP29ECO is fully controlled by the Sensistor ISH2000 Leak Detector
  • Draws a well-defined air sample to the built-in hydrogen sensor
  • Handles accumulation chamber tests, local enclosure tests and scanning
  • Suitable for hydrogen concentration monitoring
  • Designed for integration into automatic leak test systems
  • Available with two different sniffer flows – 1 cc/s or 3 cc/s

Typical Applications

  • Industry
  • Automotive
  • Aerospace
  • Packaging
  • RAC
  • Medical
  • Process

Part Number

AP29ECO Sampling Probe 3 cc/s
Part Number Description
590-035 AP29ECO Sampling Probe, 3cc/s
590-750 Sensistor ISH2000, Hydrogen Leak Detector, desktop model
590-760 Sensistor ISH2000P, Hydrogen Leak Detector, panel model
590-770 Sensistor ISH2000C, Hydrogen Leak Detector, battery model
AP29ECO Sampling Probe 1 cc/s
Part Number Description
590-036 AP29ECO Sampling Probe, 1cc/s
590-750 Sensistor ISH2000, Hydrogen Leak Detector, desktop model
590-760 Sensistor ISH2000P, Hydrogen Leak Detector, panel model
590-770 Sensistor ISH2000C, Hydrogen Leak Detector, battery model

Specifications

Type AP29ECO Sampling Probe 3 cc/s AP29ECO Sampling Probe 1 cc/s
Minimum Detectable Leak Rate 0.5 ppm H2; 3×10-5 mbarl/s or atm cc/s of 5% H2 tracer gas with standard sniffer flow 0.5 ppm H2; 1×10-5 mbarl/s or atm cc/s of 5% H2 tracer gas with standard sniffer flow
Supplies
Supply Voltage Electrical supply (24 VDC) from the Sensistor ISH2000 Hydrogen Leak Detector Fresh air with no H2 contamination Electrical supply (24 VDC) from the Sensistor ISH2000 Hydrogen Leak Detector Fresh air with no H2 contamination
Ambient temperature range 10° to 50°C (50° to 122°F) 10° to 50°C (50° to 122°F)
Dimensions (W x H x D) 92 mm x 185 mm x 260 mm (3.6 in. x 7.3 in. x 10.2 in.) 92 mm x 185 mm x 260 mm (3.6 in. x 7.3 in. x 10.2 in.)
Weight 4.2 kg (9.3 lb.) 4.2 kg (9.3 lb.)

Accessories

AP29ECO Sampling Probe 3 cc/s
Part Number Description
590-250 Insert Probe H65
591-420 Bus Cable 25 pin to pin, 2 m
590-820 Combox (incl. C21 cable 3m), to connect H65,AP29ECO, AP55 and AP57 with Sensistor ISH2000 Hydrogen Leak Detector
590-161 C21 Probe Cable, 3m (for Hand Probe P60, P60 Flex, P50, P50 Flex, H50, H51 and H21)
590-175 C21 Probe Cable, 6m (for Hand Probe P60, P60 Flex, P50, P50 Flex, H50, H51 and H21)
590-165 C21 Probe Cable, 9m (for Hand Probe P60, P60 Flex, P50, P50 Flex, H50, H51 and H21)
590-420 Calibration Leak Type A, 5x10E-2 mbarl/s @1bar
590-421 Calibration Leak Type B, 5x10E-3 mbarl/s @1bar
590-422 Calibration Leak Type C, 5x10E-4 mbarl/s @1bar
AP29ECO Sampling Probe 1 cc/s
Part Number Description
590-250 Insert Probe H65
591-420 Bus Cable 25 pin to pin, 2 m
590-820 Combox (incl. C21 cable 3m), to connect H65,AP29ECO, AP55 and AP57 with Sensistor ISH2000 Hydrogen Leak Detector
590-161 C21 Probe Cable, 3m (for Hand Probe P60, P60 Flex, P50, P50 Flex, H50, H51 and H21)
590-175 C21 Probe Cable, 6m (for Hand Probe P60, P60 Flex, P50, P50 Flex, H50, H51 and H21)
590-165 C21 Probe Cable, 9m (for Hand Probe P60, P60 Flex, P50, P50 Flex, H50, H51 and H21)
590-420 Calibration Leak Type A, 5x10E-2 mbarl/s @1bar
590-421 Calibration Leak Type B, 5x10E-3 mbarl/s @1bar
590-422 Calibration Leak Type C, 5x10E-4 mbarl/s @1bar
^