UL1000
Helium Leak Detector UL1000
Speed, Sensitivity and Reliability for Demanding Leak Detection Applications. The INFICON UL1000 Mobile Helium Leak Detector is designed to meet the most demanding industrial leak detection applications. Providing fast, accurate and repeatable test results, testing flexibility, and high sensitivity in a low maintenance design, the UL1000 is ready to tackle the toughest leak detection jobs industry can offer. The UL1000 is optimized to provide quick, accurate results in any application from the leak checking of large vessels and systems to high-cycle, repetitive component testing – 24 hours / 7 days a week – in the toughest industrial environments.
With the addition of the TC1000 Test Chamber accessory, the UL1000 Helium Leak Detector provides easy, fast and accurate testing of hermetically sealed parts like IC packages, quartz crystals and laser diodes
(according to MIL-STD 883, Method 1014).
Features
- Wide measurement range over 15 decades
- Short pumpdown and response time
- Mobile all-metal housing for added convenience with uncompromised maneuverability
- I-CAL (Intelligent Calculation Algorithm for Leak rates)to ensure fastest response time to leaks in all measurement ranges
- Zero function with automatic integration time alignment for fast and reliable leak test results
- Intelligent vacuum design with rugged roughing pump and multiple inlet turbomolecular pump that provides high helium pumping speed with high compression
- Rotatable display and user interface allows simple and easy control and interaction with the unit
- Self-protection features to protect the UL1000 from helium and particle contamination
- Auto purge cycle to ensure clean-up and readiness for leak test
- Software updates via email easily possible
- Rugged mass spectrometer system with two filament ion source (with 3-year warranty) ensures long running time and low maintenance costs
- Built-in test leak for internal calibration to ensure accurate test results
- Built-in software menu “Auto Leak Test” function to perform automatic testing of hermetically sealed components (optional TC1000 Test Chamber)
Typical Applications
- Leak testing and quality control of all types of components, including:
- automotive components
- refrigeration and air conditioning components and sub-assemblies
- hermetically sealed electronic devices
- heat exchangers
Part Number
UL1000 Helium Leak Detector | |
---|---|
Part Number | Description |
550-000A | UL1000, 230 V, EU mains plug |
550-001A | UL1000, 115 V, US mains plug |
550-002A | UL1000, 100 V AC, US mains plug |
Specifications
Type | UL1000 Helium Leak Detector | |
---|---|---|
Min. Detectable leak rate for helium ( Vacuum mode ) | mbar•L/s | <5E-12 |
Min. detectable leak rate for Helium ( Sniffer mode ) | mbar•L/s | 5E-8 |
Max. Inlet pressure GROSS mode | mbar | 15 |
Max. Inlet pressure FINE mode | mbar | 2 |
Max. Inlet pressure ULTRA mode | mbar | 0.4 |
Pumping speed during evacuation | m³/h | 16 at 50 Hz |
Helium pumping speed GROSS mode | L/s | max.8 |
Helium pumping speed FINE mode | L/s | 7 |
Helium pumping speed ULTRA mode | L/s | 2.5 |
Detectable masses | 2,3,4 amu | |
Mass spectrometer Sector field | 180° | |
Filaments ion soure Iridium/Yttria coated | 2 | |
Calibrated built in test leak | mbar•L/s | E-7 |
Test port | ||
Adjustable triggers | 2 | |
Interface | RS232 | |
Chart recorder output | V | 2×10 |
In/Outputs | PLC compatible | |
Permissable ambient temperature (during operation ) | °C | +10…..+40 |
Type of protection | IP 20 | |
Weight | kg | 110 |
Dimensions ( L x W xH ) | mm | 1068x525x850 |
Supply voltages | V (ac) | 230(+-10%)50Hz |
Power consumption | VA | 1100 |
Accessories
UL1000 Helium Leak Detector | |
---|---|
Part Number | Description |
14005 | SL200, Sniffer Probe for UL1000/5000/Modul1000, 4m length |
14022 | Extension cable for remote control RC1000, 8 m length |
14090 | LeakWare for UL1000/5000-Family |
551-000 | Toolbox, detachable,for UL1000 |
551-001 | Helium Bottle Holder for UL1000/5000 |
551-002 | ESD Mat for UL1000/5000 |
551-005 | TC1000Test Chamber,DN25KF,ESD wrist band |
551-010 | Remote Control RC1000C, wired, with 4m coiled cable |
551-015 | Remote Control RC1000WL, wireless, incl. wireless transmitter |
551-020 | Wireless transmitter for RC1000WL |