Type |
HPG400 |
Measurement range (air, N2) |
mbar |
2×10-6 … 1000 |
Accuracy |
|
10-5 … 1 mbar |
% of reading |
±15 (1) |
Repeatability |
|
10-5 … 10-1 mbar |
% of reading |
2 |
10-1 … 100 mbar |
% of reading |
30 |
Hot ion emission on, selectable: |
|
option 1 |
mbar |
1 |
option 2 |
mbar |
5×10-1 |
option 3 |
mbar |
2×10-1 |
option 4 |
mbar |
1×10-1 |
option 5 |
mbar |
5×10-2 |
Pressure, max. |
bar (absolute) |
2 |
Temperature |
|
Operation (ambient) |
°C |
0 … +50 |
Storage |
°C |
-20 … +70 |
Bakeout at flange |
°C |
80 |
Bakeout electronics removed |
°C |
150 |
Supply voltage |
VDC |
20 … 28 |
Output signal analog |
|
Output signal |
V |
0 … +10.2 |
Measurement range hot cathode |
V |
1.5 … 7.5 |
Measurement range Pirani |
V |
8.5 … 9.75 |
Voltage vs. Pressure hot cathode |
volts per decade |
1 |
Voltage vs. Pressure Pirani |
volts per decade |
0.25 |
Error signal hot cathode |
V |
0.3 |
Error signal Pirani |
V |
0.5 |
Load impedance, min. |
kΩ |
10 |
Interface (digital) (2) |
|
RS232C |
Electrical connection |
|
D-Sub, 15-pin, male |
Cable length, max. (3) |
m (ft.) |
100 (330) |
Materials exposed to vacuum |
|
Yt2O3, Ir, Pt, Mo, Cu, W, NiFe, NiCr, stainless steel, glass |
Internal volume KF / CF |
cm³ (in.³) |
20 (1.2) / 30 (1.8) |
Weight KF / CF |
g |
430 / 695 |
Degree of protection |
|
IP30 |
DeviceNet™ |
|
Protocol |
|
DeviceNet™, group 2 slave only |
Data rate switch |
kBaud |
125, 250, 500 or network programmable |
Cable length 125 kbps |
m (ft.) |
500 (1650) |
Cable length 250 kbps |
m (ft.) |
250 (825) |
Cable length 500 kbps |
m (ft.) |
100 (330) |
MAC ID |
|
2 switches (address 00 – 63) or network programmable |
Network size |
|
up to 64 nodes per segment |
Digital functions |
|
read pressure, select units: Torr, mbar, Pa; Pirani full scale adjust, monitor gauge status, safe state allows definition of behavior in case of error, detailed alarm and warning information |
Analog functions |
|
0 … 10 V analog output pressure indication, two setpoint relays A + B |
Visual communication indicators |
|
LED network status (green / red) ; LED module status (green / red) |
Specification |
|
DeviceNet™ “Vacuum Gauge Device Profile” |
Device type |
|
”CG” for combination gauge |
I / O slave messaging |
|
polling only |
Setpoint relays: Number of setpoints |
|
2 |
Setpoint relays: Range |
mbar |
2×10-6 … 100 |
Setpoint relays: Relay contact |
|
n.o., potential free |
Setpoint relays: Hysteresis |
% of reading |
10 |
Setpoint relays: Contact rating |
V (dc) |
60 |
Supply voltage for DeviceNet™ |
V (dc) |
11 … 25 |
Supply voltage for gauge |
V (dc) |
20 … 28 |
Connector for DeviceNet™ |
|
Microstyle, 5-pin |
Connector for HPG (analog output, supply voltage, setpoints) |
|
D-Sub, 15-pin, male |
Profibus DP |
|
Baud rates |
kBaud |
9.6 / 19.2 / 93.75 / 187.5 / 500 |
Address |
|
2 switches (address 00 – 127) or network programmable |
Digital functions |
|
read pressure, select units: Torr, mbar, Pa; Pirani full scale adjust, monitor gauge status, safe state allows definition of behavior in case of error, detailed alarm and warning information |
Analog functions |
|
0 … 10 V analog output pressure indication, two setpoint relays A + B |
Setpoint relays: Number of setpoints |
|
2 |
Setpoint relays: Range |
mbar |
2×10-6 … 100 |
Setpoint relays: Relay contact |
|
n.o., potential free |
Setpoint relays: Hysteresis |
% of reading |
10 |
Setpoint relays: Contact rating |
V (dc) |
60 |
Connector for Profibus DP |
|
D-Sub, 9-pin, female |
Connector for HPG (analog output, supply voltage, setpoints) |
|
D-Sub, 15-pin, male |