BPG400
ATM to Ultra-High Vacuum Gauge
The INFICON Bayard-Alpert Pirani Combination Gauge, BPG400, functions as two gauges in a single compact unit measuring from 5×10-10 mbar to atmosphere (3.8×10-10 Torr to atmosphere). Combining technologies reduces the complexity of installation, setup, and integration. Choose the BPG400 for affordable and repeatable process to base pressure measurements in one economic package.
Benefits
- Extremely wide measurement range from 5×10-10 mbar to atmosphere (3.8×10-10 Torr to atmosphere)
- Excellent repeatability in the process pressure range from 10-8 … 10-2 mbar of 5%
- The Pirani interlock protects the Bayard-Alpert system from premature filament burnout and excess contamination from high pressure operation
- Long-life yttrium oxide coated iridium filament
- Optional graphic display and Fieldbus interfaces available
- Automatic high vacuum Pirani adjustment reduces operator interventions
- RoHS compliance
- General vacuum measurement and control in the low to ultra high vacuum range
Typical Applications
- Pressure measurement in semiconductor process and transfer chambers
- Industrial coating
- General vacuum measurement and control in the low to ultra high vacuum range
Part Number
353-500 | BPG400 DN 25 ISO-KF |
---|---|
353-500-11 | BPG400 BA Pirani Gauge; DN 25 ISO-KF |
353-500-I0 | BPG400 BA Pirani Gauge; DN 25 ISO-KF |
353-502-A | BPG400 BA Pirani Gauge; DN 40 CF-R |
354-498 | Spare electronic BPG400 without Display |
354-499 | Spare electronic BPG400 with Display |
353-501 | BPG400 LCD, DN 25 ISO-KF |
353-502 | BPG400 DN 40 CF-R |
353-503 | BPG400 LCD, DN 40 CF-R |
353-505 | BPG400-SP PBus, DN 25 ISO-KF |
353-506 | BPG400-SP PBus, DN 40 CF-R |
353-507 | BPG400-SD DNet, DN 25 ISO-KF |
353-508 | BPG400-SD DNet, DN 40 CF-R |
Dimensions
Specification
Type | BPG400 | |
---|---|---|
Measurement range (air, O2, CO, N2) | mbar | 5×10-10 … 1000 |
Accuracy | ||
10-8 … 10-2 mbar | % of reading | ±15 |
Repeatability | ||
10-8 … 10-2 mbar | % of reading | 5 |
Degas (1) | ||
p < 7.2 × 10-6 | mbar | electron bombardment, max. 3 min |
Pressure, max. | bar (absolute) | 2 |
Temperature | ||
Operation (ambient) | °C | 0 … +50 |
Storage | °C | -20 … +70 |
Bakeout at flange with extension | °C | 150 |
Bakeout at flange without extension | °C | 80 |
Bakeout electronics removed | °C | 150 |
Supply voltage | VDC | +20 … +28 |
Output signal analog | ||
Output signal | V | 0 … +10 |
Measurement range | V | +0.774 … +10 |
Voltage vs. pressure | volts per decade | 0.75 |
Error signal | V | 0.3 / 0.5 |
Load impedance, min. | kΩ | 10 |
Interface (digital) (2) | RS232C | |
Electrical connection | D-Sub, 15-pin, male | |
Cable length, max. (3) | m (ft.) | 100 (330) |
Materials exposed to vacuum | Yt2O3, Ir, Pt, Mo, Cu, W, NiFe, NiCr, stainless steel, glass | |
Internal volume KF / CF | cm³ (in.³) | 24 (1.46) / 34 (2.1) |
Weight KF / CF | g | 285 / 550 |
Degree of protection | IP30 | |
Profibus DP | ||
Baud rates | kBaud | 9.6 / 19.2 / 93.75 / 187.5 / 500 |
Address | 2 switches (address 00 – 127) or network programmable | |
Digital functions | read pressure, select units: Torr, mbar, Pa; degas function, Pirani full scale adjust, monitor gauge status, safe state allows definition of behavior in case of error, detailed alarm and warning information | |
Analog functions | 0 … 10 V analog output pressure indication, two setpoint relays A + B | |
Setpoint relays: Number of setpoints | 2 | |
Setpoint relays: Range | mbar | 1×10-9 … 100 |
Setpoint relays: Relay contact | n.o., potential free | |
Setpoint relays: Hysteresis | % of reading | 10 |
Setpoint relays: Contact rating | VDC | 60 |
Connector for Profibus DP | D-Sub, 9-pin, female | |
Connector for BPG (analog output, supply voltage, setpoints) | D-Sub, 15-pin, male | |
DeviceNet™ | ||
Protocol | DeviceNet™, group 2 slave only | |
Data rate switch | kBaud | 125, 250, 500 or network programmable |
Cable length 125 kbps | m (ft.) | 500 (1650) |
Cable length 250 kbps | m (ft.) | 250 (825) |
Cable length 500 kbps | m (ft.) | 100 (330) |
MAC ID | 2 switches (address 00 – 63) or network programmable | |
Digital functions | read pressure, select units: Torr, mbar, Pa; degas function, Pirani full scale adjust, monitor gauge status, safe state allows definition of behavior in case of error, detailed alarm and warning information | |
Analog functions | 0 … 10 V analog output pressure indication, two setpoint relays A + B | |
Specification | DeviceNet™ “Vacuum Gauge Device Profile” | |
Device type | ”CG” for combination gauge | |
I / O slave messaging | polling only | |
Setpoint relays: Number of setpoints | 2 | |
Setpoint relays: Range | mbar | 1×10-9 … 100 |
Setpoint relays: Relay contact | n.o., potential free | |
Setpoint relays: Hysteresis | % of reading | 10 |
Setpoint relays: Contact rating | VDC | 60 |
Supply voltage for DeviceNet™ | VDC | +11 … +25 |
Supply voltage for gauge | VDC | +20 … +28 |
Connector for DeviceNet™ | Microstyle, 5-pin | |
Connector for BPG (analog output, supply voltage, setpoints) | D-Sub, 15-pin, male |
1) Reduced accuracy during degas
2) Simultaneous use of RS232C or VGC40x controller series and Fieldbus is not allowed
3) For RS232C operation <30m
Accessories
Part Number | Description |
---|---|
353-510 | 100 mm sensor extension tube |
353-512 | Baffle for BPG400 |
211-113 | Centering ring with baffle DN 25 KF |
Spare Parts
Part Number | Description |
---|---|
354-490 | Spare sensor head BPG400, DN 25 ISO-KF |
354-491 | Spare sensor head BPG400, DN 40 CF-R |