your best partner for vacuum

UL6000 FAB

The ultimate in mobile vacuum leak detection

The INFICON UL6000 Fab Product Seies is designed to meet the most critical and demanding leak detection applications. Featuring INFICON proprietary software algorithms I·CAL and HYDRO·S in a field-proven vacuum design, the UL6000 provides testing flexibility, high sensitivity and quick accurate results making any leak detection application fast and easy.

The UL6000 delivers fastest response times in all measurement ranges and extremely short cycle times in reaching test conditions and final results. The specially designed vacuum architecture provides the continuous high helium pumping speeds and the extremely fast response times you demand.

The all new UL6000 Fab PLUS with I·RISE. Rate-of-rise in under 10 seconds

Leak detection is both time-consuming and complex. By measuring rate-of-rise, semiconductor- and vacuum equipment is continuously monitored during operation. Upon leak indication by this integral leak test, leak positions are then localized using the helium spraying method allowing for the required repair. This is followed by another leak test using pressure increase or so-called vacuum decay measurement – an intricate process which needs to be repeated until all leaks are detected. Moreover, to be as precise as possible, measurement times for this procedure are often particularly long for large equipment volumes. Thanks to I·RISE, our Intelligent Rate-of-rise feature, the new UL6000 Fab PLUS now performs pressure increase measurement all by itself and combines it with the helium spraying method. Our patented system enables measurement times to drop drastically, and thus helps minimize both general time for leak testing and fab equipment maintenance. Experience even faster and more precise measurement results – in under 10 seconds, independent of chamber volume.

Features

  • UL6000 Fab Product Series – Advantages
    • ULTRATEST Sensor Technology
    • HYDRO-S (HYDROgen-Suppression) to enable test conditions to be reached quickly.
    • I-CAL (Intelligent – Calculation Algorithm for Leak Rates) to ensure fast response to leaks in all measurement ranges.
    • A special multiple inlet turbomolecular pump that provides optimized high helium pumping speeds with high helium compression at all pressures to deliver excellent sensitivity, response, and fast clean-up.
    • A booster turbomolecular pump for extremely fast response on any test object, including the largest volumes and chambers.
    • The field-proven INFICON magnetic sector mass spectrometer for years of accurate and repeatable results in a trouble-free, maintenance-free design.
    • Self-protection features to protect the UL6000 from helium and particulate contamination and even an auto-purge cycle to ensure the UL6000 is ready when you need it.
    • A rotatable display / control interface and optional remote hand controller for operational flexibility.
    • Access control to the unit and system software can be enabled via the software to prevent unauthorized use or unintended modifications to test set-ups.
    • Software updates via e-mail to keep your UL6000 at optimum performance.
    • Work Station Design with an optimal height work surface that includes an ESD mat and tool box for testing convenience.
  • UL6000 Fab PLUS – Additional Advantages
    • Featuring I·RISE, our intelligent rate-of-rise: Provides in-device pressure increase testing combined with helium spraying method
    • Rate-of-rise measurements in under 10 seconds
    • Helps minimize general time for leak testing and fab equipment maintenance
    • More speed, more precision

Typical Applications

  • Leak testing of components, subassemblies and large chambers used on semiconductor and flat panel display tools
  • Leak testing of semiconductor and flat panel tools
  • Leak testing of aerospace components, subassemblies, and systems
  • Leak testing of storage vessels/storage tanks

Part Number

UL6000 Fab
Part Number Description
550-520 UL6000 Fab – mobile leak detector
UL6000 Fab PLUS
Part Number Description
550-530 UL6000 Fab PLUS – mobile leak detector

Specifications

Min. detectable leak rate for Helium (vacuum mode) mbar•L/s < 5 • 10-12 < 5 • 10-12
Min. detectable leak rate for Helium (sniffer mode) mbar•L/s < 5 • 10-8 < 5 • 10-8
Max. pumping speed of roughing pump m³/h >36 >36
Max. Helium pumping speed (ULTRA Mode) L/s >36 >36
Response Time for 50 L s 1.4 1.4
Venting time 50 L to atm. s 16 16
Evacuation Time to 1 mbar 50 L s 52 52
Run up time for operation s <2 min. <2 min.
Calibration time s <30 <30
Time for Rate-of-rise measurement (I·RISE) s <10
Inlet flange DN63 ISO-K DN63 ISO-K
Detectable masses (ULTRATEST Sensor Technology) 2, 3, 4 (H23He, He) 2, 3, 4 (H23He, He)
Supply voltage 100 to 240 V 50/60 Hz (AC) 100 to 240 V 50/60 Hz (AC)
Max. Power consumption VA 1500 1500
Typ. Power consumption VA 700 700
Dimensions (LxWxH) 42x21x41 42x21x41
Weight kg 143 144

Accessories

UL6000 Fab
Part Number Description
551-226 Metal hose, DN63 ISO-K/DN50 ISO-KF
551-227 Metal hose, DN63 ISO-K/DN40 ISO-KF
551-015 Remote Control RC1000WL, wireless
551-010 Remote Control RC1000C, with 4m cable
551-202 Vaccuum Hose Holder for UL3000 Fab
525-001 SL3000-3,sniffer line 3m length
525-002 SL3000-5,sniffer line 5m length
525-003 SL3000-10, sniffer line 10m length
551-203 Bracket for SL3000
551-210 SL200, Sniffer line + ext. for UL3000
551-201 Helium Bottle Holder for UL3000/6000 Fab
560-310 I/O 1000 module
560-315 BM1000 Profibus module
560-316 BM1000 PROFINET IO
560-317 BM1000 DeviceNet
560-318 BM1000 EtherNet/IP(TM)
560-334 Data Cable for IO1000,BM1000, 0.5m
560-332 Data cable 2 m LDS3000
560-335 Data Cable 5 m
560-340 Data Cable 10 m

Consumables

Part Number Description
200007656 Airfilter Kit UL3 Housing Ventilation

Datasheets

^